CHARACTERIZATION OF DELUX: ULTRAVIOLET LIGHT
STERILIZATION DEVICE FOR PFF2/ N95 MASKS AGAINST COVID-19 |
||
CARACTERIZACIÓN DO DELUX: DISPOSITIVO DE ESTERILIZACIÓN POR LUZ ULTRAVIOLETA PARA MÁSCARAS PFF2/ N95 CONTRA COVID-19 |
Christiane Bertachini Lombello1,* , Nasser Ali Daghastanli1 , Hermann Windisch Neto1, Fernando Luiz Affonso Fonseca 2, Felipe Nogueira Ambrosio1 , Raquel Arbex1 , |
Received: 15-11-2021, Received after review:
20-12-2021, Accepted: 26-12-2021, Published: 01-01-2022 |
Abstract |
Resumen |
The World Health Organization (WHO) has
declared a public health pandemic state due to the transmission of the new
coronavirus on March 11th, 2020. COVID-19, that is caused by SARS-CoV-2, has
a very broad clinical spectrum, with predominantly respiratory symptoms
developments. The role of the health professionals in fighting the pandemic
requires the use of Personal Protective Equipment (PPE). PFF2 / N95 masks are
suitable PPEs for this purpose. Due to the high demand for PFF2 / N95 masks
to fight the pandemic, there was a shortage of this PPE worldwide. This work
aims to present a characterization of the device called DELUX, utilized for
UVC (ultraviolet C light spectra) sterilization of PFF2 / N95 masks, and
allowing to extend the time of safe use of this PPE in emergency conditions.
The photometric validation of the device resulted in the verification of the
emission spectrum of the lamps used in the device, and the measurement of the
optical power, demonstrating the adequacy of irradiation with UVC light, with
a 15-minute cycle, and safely. Biological validation showed that DELUX is
capable of inactivating SARS-CoV-2 present on the surface of PFF2 / N95
masks, thus being efficient for their sterilization. The safety offered by
the sterilization cycle allows to extend the safe use of those masks. |
La Organización Mundial de la Salud (OMS) declaró un estado pandémico de salud pública debido a transmisión del nuevo coronavirus el 11 de marzo de 2020. La COVID-19, causada por el SARS-CoV-2, tiene un espectro clínico muy amplio, con predomínio de desarrollos sintomáticos respiratorios. El papel de los profesionales de salud en la lucha contra pandemia requiere el uso de Equipo de Protección Personal (EPP). Las máscaras PFF2 / N95 son EPP adecuados para este propósito. Debido a gran demanda de máscaras PFF2 / N95 para combatir la pandemia, hubo escasez en el mercado. Este trabajo tiene como objetivo caracterizar un dispositivo denominado DELUX para la esterilización por UVC (espectros de luz ultravioleta C) de mascarillas PFF2 / N95, ampliando el tiempo de uso seguro de este EPP en condiciones de emergencia. La validación fotométrica del dispositivo permitió la verificación del espectro de emisión de las lámparas utilizadas en el dispositivo y la medición de la potencia óptica, demostrando la idoneidad de la irradiación con luz UVC, en ciclos de 15 minutos, de forma segura. La validación biológica mostró que DELUX es capaz de inactivar el SARS-CoV-2 presente en la superficie de las mascarillas PFF2 / N95, siendo así eficiente para su esterilización. La seguridad que ofrece el ciclo de esterilización permite extender el tiempo de uso seguro de estas mascarillas. |
Keywords: COVID-19, equipment, protective,
radiation, sterilization, ultraviolet |
Palabras clave: COVID-19, equipo, esterilización, protección, radiación, ultravioleta |
1,*CECS, Federal University of
ABC, UFABC, Brazil. Corresponding author ✉: christiane.lombello@ufabc.edu.br. 2ABC Medical School,
FMABC, Brazil Suggested citation: Lombello, C. B.; Ali Daghastanli, N:; Windisch Neto, H.; Affonso Fonseca, F. L.; Nogueira Ambrosio, F.; Arbex, R.; Kobb Fernando, I.; da Silva, V. A.; Pataquini dos Santos, L. and Aparecida da Ana, P. “Characterization of DELUX: Ultraviolet light sterilization device for PFF2 / N95 masks against COVID-19,” Ingenius, Revista de Ciencia y Tecnología, N.◦ 27, pp. 23-31, 2022, doi: https://doi.org/10.17163/ings.n27.2022.03 |
1.
Introduction Viruses known as coronaviruses
belong to the order Nidovirales, family Coronaviridae, subfamily Coronavirinae,
genus Coronavirus (CoVs) [1]. They were first
identified in poultry in the 1930s and many of them have been associated with
respiratory, gastrointestinal, liver and neurological diseases in animals.
Some coronaviruses are known to cause disease in humans [2]. Of these, four
species (229E, OC43, NL63 and HUK1) usually cause symptoms of a common cold
in humans, and only rarely can result on a severe lower respiratory tract
infection, such as pneumonia. These cases are particularly important in
children, the elderly and immunodepressed patients
[3–5]. However, three other
species considered zoonoses (SARS-CoV, MERS-CoV and SARS-CoV-2) are
associated with more severe respiratory infections in humans, and can be
fatal [3,4]. SARS-CoV was identified in 2002 as the
cause of Severe Acute Respiratory Syndrome (SARS), while MERS-CoV was identified in 2012 as the cause of Middle Eastern
Respiratory Syndrome (MERS) [4]. More recently, SARS-CoV-2 (Figure 1), known
as the new coronavirus, was first detected on December 2019 in the city of
Wuhan (China), and identified as the etiologic agent of the disease capable
of aggravating the respiratory system [3], [6] with an important impact on
global public health due to consequent morbidity and mortality worldwide [7,
8]. According to the World Health Organization (WHO), the official
nomenclature for the disease caused by this virus was defined as Coronavirus
Disease-2019 (COVID-19) and, on March 11, 2020, WHO declared a state of
pandemic in public health, due to the transmission of the new coronavirus [9,
10]. COVID-19 disease,
caused by SARS-CoV-2, has a very broad clinical spectrum. There are
asymptomatic variations or diverse clinical symptoms, such as sore throat,
diarrhea, anosmia or hyposmia, myalgia, tiredness,
fatigue, and skin manifestations, like dermatoses and urticaria
were also recorded [8], [11, 12]. However, mainly respiratory symptoms such as
dry cough and shortness of breath, associated with fever, and other symptoms
mentioned above are directly related to the high rates of hospitalization and
death resulting from the disease [8], [12]. The health
professionals work is fundamental during the infection combat, in prevention,
detection, patient treatment and recovery, and these professional efforts are
recognized worldwide [13]. During the professional activities the use of
Personal Protective Equipment (PPE) can provide safety, preventing accidents
or occupational diseases. The Brazilian Health Regulatory Agency (Agência Nacional de Vigilância Sanitária, Anvisa) determines
that the PPE required during the activities performed by health professionals
with possible exposition to SARS-CoV-2 include: cap, protection |
glasses or face shield, mask
(surgical or respiratory protection mask), long-sleeved coat and gloves [14].
Figure
1.
Structure of the coronavirus. The envelope, capsid proteins (S and HE), N
protein and viral RNA are represented As
this is a disease with preferential transmission through the air, masks are
essential PPE in controlling the spread of the disease. These devices are
also called Respiratory Protection Equipment (RPE). Surgical masks are
indicated to prevent the contamination of the professionals’ airways by
respiratory droplets (particles larger than 5 µm) when acting less than 1
meter from the patient. However, not all the masks provide protection against
aerosols [14,15]. Although these masks, as well as fabric masks, are not
considered PPE by Anvisa, their use by the
population was mandatory in several countries, like Brazilian one, in order
to contain the transmission of COVID-19. In health services, masks must be
used by asymptomatic patients, visitors, companions and professionals in
administrative areas who do not have contact with patients at distances of
less than 1 meter [14], [16]. The
masks called semi-facial filtering piece (PFF) are PPEs with a minimum
efficiency of 95% in the filtration of particles up to 0.3 µm. The
classification of masks occurs according to the level of penetration and
breathing resistance, which can be PFF1, PFF2 and PFF3 in Brazil. For the
protection of aerosols (that is, particles smaller than droplets) containing
biological agents, PFF2 must be used, equivalent to the N95 mask adopted in
the United States [14], [16–18]. These
RPEs should be used by healthcare professionals who work in procedures with
risk of generating aerosols, such as collection of nasotracheal
secretions, bronchoscopy, non-invasive ventilation, manual ventilation before
intubation, intubation or tracheal aspiration, and cardiopulmonary
resuscitation. In addition, they are also recommended to support
professionals who develop their activities in areas of carrying out procedures
that generate aerosols [14], [16]. However, in cases of prolonged use, it is
important that this equipment remains adjusted in the face, and functional.
In this context, the Brazilian Institute for Patient Safety (IBSP) warns that
these masks remain effective when used for up to 8 consecutive hours [19].
PFF2/N95 masks are formed by superimposed layers of fabricnon-woven
(TNT) polypropylene, which filter the air |
and trap particles in the order of
0.3 µm, including microorganisms such as the SARS-CoV-2 virus [20]. The impact of the
COVID-19 pandemic on healthcare services included the high demand for PPE
used by healthcare professionals in the combat of the coronavirus, resulting
in a shortage of supply of this equipment. Although not ideal, the lack of
availability of PFF2/N95 masks on the market makes the reuse of these PPEs
necessary. The Anvisa angency
states that respiratory protection masks may, exceptionally, be used for a
longer period or for several times greater than that foreseen by the manufacturer.
However, the reuse of these PPE for a time beyond what is considered safe
exposes health professionals to contamination by the virus, due to the
conditions of prolonged use, handling and storage. Some Anvisa
recommendations described in the Technical Note GVIMS/GGTES/ANVISA N. 04/2020
[16] indicates the proper use of this PPE that must be followed. The
reprocessing of a PPE regularized at Anvisa as
disposable must ensure that it is as safe as a new PPE, in addition to not
affecting the performance characteristics. Possible reprocessing protocols
must be validated in order to guarantee the sterilization of the product as
well as its integrity and functionality [16]. Among the
alternatives developed to ensure safety in the extension of the useful life
and reuse of these masks, the sterilization of these equipments
with UVC radiation has shown to be promising. This technique, in addition to
enabling the reuse of masks quickly, does not require large infrastructure
and elaborate spaces, as well as the frequent replacement of supplies. Thus,
the use of a sterilization device with such a function guarantees a versatile
possibility so that, at first, healthcare workers can sterilize their PPE
safely and quickly [21]. This paper aims to
characterize a low-cost and fast production equipment, developed in the ABC
region of São Paulo, for the sterilization of PFF2/N95 masks, with the
proposal to extend the time of use of this PPE in emergency conditions of
shortages in the market due to the current COVID-19 pandemic. 2. Materials and methods 2.1. Emission of UVC lamps The characterizations of the
emission spectrum of the UVC lamps (PURITEC 15W, Osram,
Brasil) used in the DELUX device was carried out
using three different devices. 2.1.1.
Ultraviolet-visible
photodiode spectrophotometer (Cary 50, Varian) Measurements were taken to obtain
the emission spectrum of UVC lamps (PURITEC 15W, Osram,
Brasil). |
2.1.2.
OCEAN OPTICS
spectrometer (UV- VIS- NIR spectrometer) This spectrometer is
coupled to an optical fiber (φ=400 nm) to measure the emission of the UVC lamp in loco, that is,
inside the sterilization and measurement equipment of the internal
environment of the irradiation chamber. 2.1.3.
FieldMaxII-TO optical powermeter (Coherent- USA) Powermeter was coupled to the PM10 sensor
(Coherent-USA) to measure the power values per area (mW/cm2)
through the PM10 semiconductor sensor in the desired spectral region. 2.2. Biological Tests Validation of
efficacy was carried out together with the ABC Medical School (FMABC). To
ensure the correct PFF2/N95 mask sterilization through the DELUX system, a
methodology was proposed to assess the effectiveness against the presence of
viruses after the sterilization cycle. For this study the virus were detected using RT-PCR (polymerase chain reaction)
[22, 23]. Three
N95 masks of the same type used by health professionals during their daily
activities were used (in particular the respiratory mask 8801H 9 PFF-2
Respirators for biological risks (Brand: 3M®, Brasil)
A control sample with the SARS-CoV-2 virus was used to mimic the
contamination on the front of the mask, following the experimental
procedures. Initially a swab containing a viral sample was used to spread the
contents from bottom to top and left to right. The swab was used for 6 times
for spreading in the anterior direction of each of the masks, and rested at
room temperature for 3 minutes, in biosafety cabinet. The masks were then
positioned in the DELUX device in order to maximize their complete
illumination with UVC light. Irradiation of the masks with UVC light was
carried out in a single 15-minute cycle. After
irradiation, samples were collected from the masks. For this purpose, the
masks were kept for 1 minute in a biosafety cabinet, and using a sterile swab
was rubbed on the surface of the mask in the anterior region from the bottom
to the top and from the left to the right, repeated 6 times. The swab was
placed in an extraction solution in order to obtain the viral RNA. RNA
extraction and amplification followed the Centers for Disease Control and
Prevention (CDC) protocol. 3.
Results and
discussion For the
sterilization of PFF2/N95 masks, one of the possible measures adopted worldwide
is sterilization |
with UVC radiation, a physical
agent. Ultraviolet light is in the electromagnetic spectrum, and UVC
radiation comprises wavelengths between 190 and 280 nm, located within the
spectrum of UV light (electromagnetic spectrum between 100 and 390 nm).
However, wavelengths below 200 nm have no biological significance, as they
are intensely absorbed by air [24]. UVC radiation lamps
used for germicidal action usually have peak emission at a wavelength of 254
nm. This parameter is suitable for the inactivation of several
microorganisms, including coronaviruses, as stated in previous studies with
SARS-CoV and MERS-CoV
[25, 26]. The photolytic effect of UVC radiation is capable of destroying or
inactivating the microorganism, preventing it from multiplying. It is
intensely absorbed by nucleic acids and proteins, causing the disruption of
nucleic acid bases and the inactivation of enzymes. Unlike other techniques,
its photolytic action rarely produces potentially dangerous by products [27–29]. The device called
DELUX (Figure 2) was developed in partnerships with the ABC School of
Medicine, FMABC, and with the company based in the ABC region of São Paulo
(named Ecosan). The device was made with materials
compatible with the purpose and use, mainly using 304 stainless steel sheets,
carrying handle, electrical system to drive 4 ultraviolet light lamps
(PURITEC 15W, Osram, Brasil),
automated timer system, lamp activation indicator LED and safety system to
prevent the activation of the lamps with the equipment open.
(a)
(b) Figure 2. DELUX. (a) prototype, (b)
device The
device allows the simultaneous accommodation of up to 16 masks, in niches
that allow the identification of each mask. It was carried out with the
validation of the photometric efficacy of the device and the verification of
the biological efficacy against the coronavirus. |
3.1. Photometric Measures Ultraviolet light
(UV) is part of the electromagnetic spectrum characterized by wavelengths (λ) smaller than those of the
visible light (VIS) spectrum, which varies between 400nm - 800nm. The
spectral range of UV light is between 100 - 400 nm, and it is divided into 3
different regions: UVA (long-wave) between 315 - 400 nm; UVB (medium-wave)
between 280-315 nm, and UVC (short-wave) between 100 - 280 nm [24]. UVC light
has a germicidal effect compared to visible light [24], [28, 29]. Regarding
safety, exposure to UVC can cause several adverse biological effects, such as
acute inflammation of human tissues, conjunctivitis, erythema, in addition to
being associated with chromosomal alterations that can even cause some types
of cancer [24]. Thus, the use of UV light must be carried out carefully to
prevent possible adverse effects. The biological effects of the action of UV
light can be evaluated as a function of damage through the parameter of
Relative Spectral Effectiveness, S(λ), thus when exposure to UV light occurs,
care must be taken [24], [26], [28, 29]. The measurements of the optical power of the lamps used in the DELUX prototype were started with the photometric characterization of the PURITEC 15W, (Osram, Brasil), by measuring the emission spectrum in the UV-VIS region (Varian spectrophotometer), as shown in Figure 3. Typical emission peaks at specific wavelengths were observed.
(a)
(b) Figure 3. (a) Experimental setup with the light probe (Ocean Optics). (b) Emission spectra of a white light lamp (LampTETO) and UVC light (LampUV). Emission peaks in the region of interest, below 400 nm. |
The measurements of
the power emission intensities were complemented with the optical powermeter. Measurements of the power emitted by the UVC
lamp were performed using peak emission at λ = 256 nm, and λ = 438 nm, shown in Figure 4. These data
allowed us to evaluate the emission of the UVC lamps used in DELUX at the
relevant wavelengths (λ) (Figure 4).
Figure 4. Assembly for
measuring the power of UV light emitted by UV lamps. Measurements were
performed with two different power probes (Coherent), OP2-VIS and PM10 (for
UVC). Height (H), optical power (P) Temporal
measurements of UVC light emission (at 256 nm) were performed from the moment
the lamps were turned on at an interval of 800 s. Figure 4 shows the kinetic
curve of the UV lamp emission. We observe that the power emission efficiency
of the lamp increases from exponential-like transient to constant emission
(Figure 5). Measurements were
performed using the UV lamps in the interior of the device, and possible leak
of light to the exterior of the device was verified. The equipment has the
installation of rubber blankets that were effective for the complete blocking
of light. Measurements of light power were not obtained on the outside of the
device, closed and turned on, with activation of the lamps, which is an
important result to guarantee safety for users. For sterilization to
be effective, some factors must be taken into account. The dose (or intensity)
of light delivered (J/cm2) is determined by the product between
the irradiance of the light source (mW/cm2)
and the irradiation time. The dose of light must be sufficient and adequate
to penetrate the material to be sterilized. The penetration of light
radiation is controlled by the material’s absorption coefficient: the lower
the coefficient, the greater the success of the application [28, 29]. |
Figure 5. Kinetic curve of
UV lamp emission. Temporal measurements. Red dots are the monoexponential
fit of light emission at 256 nm (10Hz). Black line is dot smoothing (FFT -
Origin). Solid green line is the monoexponential
fit of the experimental data, UVC lamp emission reaches maximum value of 25mW
after 400s. Minimum power emitted 8 mW The sterilization
efficiency also depends on the regularity and exposed area on the material
surface, as there is a minimum dose of UVC radiation to inactivate the
microorganisms [20,21], [28,29]. Thus, a successful sterilization will only
occur if the entire surface of the material is exposed to UVC radiation, as
any shadows produced by the fabric layers of the masks can reduce the
effectiveness of the sterilization. As the surface requires high intensity UV
light, the lamps must be close to the material to be sterilized [30]. Different mask
models may have different sterilization results [21]. Models with extra
layers of outer protection can make it difficult or even prevent UVC light
from penetrating the innermost layers. There are more studies that allow the
standardization of this procedure, and individual tests are recommended for
each type of mask that is used in a particular location [21], [31]. The recommended dose
of UVC radiation to ensure effective sterilization is 1 J/cm2. The
dose between 0.5- 1.8 J/cm2 was able to inactivate 99.9% of pathogens, such
as some types of Influenza A, MERS-CoV, SARS-CoV and the MS2 bacteriophage. The mask’s filtering
capacity was maintained even with high doses of UVC radiation [20, 21],
[28–31]. The use of UVC light
is a possible solution for the sterilization of N95 masks in their reuse, as
they have antimicrobial and antiviral effects and do not generate toxic
residues that may remain in the masks. UVC radiation can promote the
degradation of the polypropylene that makes up the filtering layers, but this
is a slow process, in which doses lower than 950 J/cm2 do not
present any significant loss of filtration efficiency [32, 33]. |
3.2. Biological Test Upon analyzing the results, all
samples were classified as «Not Detected». It was possible to conclude that
the samples tested after the SARS-CoV-2 scattering procedure and submitted to
DELUX presented a «Not Detected» result, showing that the 15 minute UVC light
cycle with the DELUX device was sufficient to sterilize the masks. Due to high demand
associated with production and logistical difficulties, the quantity of N95
masks was shown to be limited and insufficient to attend healthcare
professionals during a pandemic situation [34, 35]. The Anvisa
determines that, due to the public health emergency situations,
exceptionally, these masks can be used for longer periods of time and for a
several times greater than the specifications provided by the manufacturer,
which must be followed by some precautions. These must include PPE
protection, proper handling and inspection before each period of usage [36,
37]. The reuse limit is not stipulated by Anvisa,
which must be recommended by each healthcare facility, and proper protocols
related to the reuse of these PPE are encouraged, as well as, health
professionals care with the PPE. A review study
demonstrated that an irradiation dose of 4 J/cm2 resulted in a 3-log
(99.9%) reduction in viral presence, which represents a total decontamination
in models using Influenza virus [31]. In addition to the irradiation dose,
the time elapsed after contamination also results in a decrease in the
presence of microorganisms, with a reduction of 1 log (90%) occurring after
24 hours [31]. Studies consider a period of 60 seconds of irradiation for
sterilization [38]. For the MERS-CoV virus, the
appropriate time was 5 minutes [26]. Meanwhile for the H1N1 virus, the
appropriate time was 15 minutes, and the distance between the UVC lamp and
the masks was 25 centimeters [39]. The DELUX device was based on 15 minute cycle sterilization, proposing a unique cycle
per day per mask. In Brazil, due to
lack of founds, many centers do not have space or
equipment to implement or develop protocols regarding sterilization by
ultraviolet radiation (UVC), and the lack of this PPE has been a persistent
problem since the beginning of the pandemic. This study allowed the
characterization of the DELUX device that can be used to sterilize PFF2/N95
masks using UVC radiation, in a standardized way, ensuring the extension of
the time of use of these PPEs safely for health professionals, as shown by
photometric and biological characterizations. Another advantage of
this UVC PFF2/N95 masks sterilization includes the reduction of potentially
infectious waste generated by the disposal of the large amount of masks used
in the COVID-19 pandemic [40]. |
4.
Conclusions The DELUX device is based on the
UVC sterilization strategy for irradiating PFF2/N95 masks with a 15 minute cycle. It is a low-cost, fast and effective
device to sterilize PFF2/N95 masks, allowing for its proper and safe reuse in
order to alleviate the shortage of PPE during the COVID-19 pandemic. The
photometric measurements show that the DELUX can make an adequate UVC-light
irradiation, resulting in an efficient sterilization on masks. Additionaly, photometric measurements allowed,
guarantying a safe device to handle with. The biologic characterization
resulted in inactivation of SARS-CoV-2 on the masks surface. Acknowledgments The UFBAC for finantial
support, UFABC Multiuser Center (CEM UFABC) for the equipment support, the
ABC Medical School (FMABC) for the partnership and biological validation, and
Ecosan for the development of the DELUX device. References [1] S. Payne,
“Family coronaviridae,” Elsevier Public Health
Emergency Collection, pp. 149–158, 2017. [Online]. Available:
https://dx.doi.org/10.1016/ B978-0-12-803109-4.00017-9 [2] J. S. M. Peiris, “Coronaviruses,” Elsevier Public Health
Emergency Collection, pp. 587–593. [Online]. Available:
https://dx.doi.org/10.1016/ B978-0-7020-4089-4.00072-X [3] M. D. S. Cespedes and J. C. R. P. Souza, “Sars-cov-2: A clinical
update - ii,” Revista da Associação Médica Brasileira, vol. 66, no. 4, pp. 547–557. [Online].
Available: https://doi.org/10.1590/1806-9282.66.4.547 [4] M. Pal, G. Berhanu, C. Desalegn, and V. Kandi, “Severe Acute Respiratory Syndrome Coronavirus-2
(SARS-CoV-2): An Update,” Pal M, Berhanu G, Desalegn C, Kandi V. Severe
Acute Respiratory Syndrome Coronavirus-2 (SARS-CoV-2): An Update. Cureus. 2020;12(3): e7423. Published 2020 Mar 26. doi:10.7759/cureus.7423, vol. 12,
no. 3, p. e7423, 2020. [Online].
Available: https://dx.doi.org/10.7759/cureus.7423 [5] D. X. Liu, J. Q. Liang, and T. S. Fung, “Human coronavirus-229e, -oc43, -nl63, and -hku1 (coronaviridae),” Elsevier Public Health Emergency Collection, pp. 428–440. [Online]. Available: https://dx.doi.org/10.1016/ B978-0-12-809633-8.21501-X
|
[6] M. Wang, R. Cao,
L. Zhang, X. Yang, J. Liu, M. Xu, Z. Shi, Z. Hu, W. Zhong,
and G. Xiao, “Remdesivir and chloroquine
effectively inhibit the recently emerged novel coronavirus (2019-nCoV) in
vitro,” Cell research, vol. 30, no. 3, pp. 269–271, 2020. [Online].
Available: https://doi.org/10.34172/jrcm.2020.005 [7] S. H. Alfaraj, J. A. Al-Tawfiq, A. Y.
Assiri, N. A. Alzahrani,
A. A. Alanazi, and Z. A. Memish,
“Clinical predictors of mortality of Middle East Respiratory Syndrome
Coronavirus (MERS-CoV) infection: A cohort study,” Travel
Medicine and Infectious Disease, vol. 29, pp. 48–50. [Online]. Available:
https://doi.org/10.1016/j.tmaid.2019.03.004 [8] L. Piroth, J. Cottenet, A.-S. Mariet, P. Bonniaud, M. Blot, P. Tubert-Bitter, and C. Quantin, “Comparison of the characteristics, morbidity, and mortality of COVID-19 and seasonal influenza: a nationwide, population-based retrospective cohort study,” The Lancet Respiratory Medicine, vol. 9, no. 3, pp. 251–259, 2021. [Online]. Available: https://doi.org/10.1016/S2213-2600(20)30527-0 [9] FIOCRUZ. (2021) Por que a doença causada pelo novo coronavírus recebeu o nome de Covid-19? Fundação Oswaldo Cruz. [Online]. Available: https://bit.ly/3pwoiEE [10] J. Zhang, W. Wu, X. Zhao, and W. Zhang, “Recommended psychological crisis intervention response to the 2019 novel coronavirus pneumonia outbreak in China: a model of West China Hospital,” Precision Clinical Medicine, vol. 3, no. 1, pp. 3–8, 02 2020. [Online]. Available: https://doi.org/10.1093/pcmedi/pbaa006 [11] M. Relvas, J. C. ao, R. Oliveira, J.
C. Cardoso, and M. Gonçalo, “Manifestações
cutaneas associadas a
COVID-19: Uma revisão
narrativa,” Revista Científica da Ordem dos
Médicos, vol. 34, no. 2, pp. 128–136, 2021. [Online]. Available: https://doi.org/10.20344/amp.14574 [12] M. C. Grant, L. Geoghegan, M. Arbyn, Z. Mohammed, L. McGuinness, E. L. Clarke, and R. G. Wade, “The prevalence of symptoms in 24,410 adults infected by the novel coronavirus (SARS-CoV-2; COVID-19): A systematic review and meta-analysis of 148 studies from 9 countries,” PLoS One, vol. 15, no. 6, p. e0234765, 2020. [Online]. Available: https://doi.org/10.1371/journal.pone.0234765 [13] E. A. S. Medeiros, “A luta dos profissionais de sade no enfrentamento da COVID-19,” Acta Paul Enferm., p. 33, 2020. [Online]. Available: http: //dx.doi.org/10.37689/acta-ape/2020EDT0003
|
[14] Secretaria de Vigiláncia em Saúde. (2021) Recomendações de proteção aos trabalhadores do serviço de saúde no contexto da
pandemia da Covid-19. Ministério da Saúde, Brasil.
[Online]. Available: https://bit.ly/3pvZ3lX [15] C. D. Cappa, S. B. Sima Asadi, A. S. Wexler, N. M. Bouvier, and W. D. Ristenpart, “Expiratory aerosol particle escape from surgical masks due to imperfect sealing,” Scientific Reports, no. 11, p. 12110, 2021. [Online]. Available: https://doi.org/10.1038/s41598-021-91487-7 [16] ANVISA, Nota Técnica GVIMS/GGTES/AN-VISA N◦ 04/2020. Orientações para serviços de saúde: medidas de prevenção e controle que devem ser adotadas durante a assistencia aos casos suspeitos ou confirmados de infecção pelo novo coronavirus (SARS-CoV-2). Agéncia Nacional de Vigiláncia Sanitária, 2021. [Online]. Available: https://bit.ly/3z12T9X [17] ABNT, Equipamento de proteção respiratória – Filtros para partículas. Associação Brasileira de Normas Técnicas (ABNT), 2010. [Online]. Available: https://bit.ly/3qwTiUF [18] ——, Equipamento de proteção respiratória – Peça semifacial filtrante para partículas. Associação Brasileira de Normas Técnicas (ABNT). [Online]. Available: https://bit.ly/3qqm09z [19] IBSP. (2020) Máscaras n95 – recomenda˜ccões para uso prolongado e reutiliza˜ccão. Instituto Brasileiro Para Segurança Do Paciente. [Online]. Available: https://bit.ly/3Jpfd8D [20] S.
Duncan, P. Bodurtha, and S. Naqvi, “The protective
performance of reusable cloth face masks, disposable
procedure masks, kn95 masks and n95 respirators: Filtration and total inward
leakage,” PLoS One, vol. 16, no. 10,
p. e0258191, 2021. [Online]. Available: https://doi.org/10.1371/journal.pone.0258191 [21] C.
C. Ontiveros, D. C. Shoults, S. MacIsaac,
K. D. Rauch, C. L. Sweeney, A. K. Stoddart, and G.
A. Gagnon, “Specificity of UV-C LED disinfection efficacy for three N95
respirators,” Scientific Reports, no. 11, p. 15350, 2021. [Online].
Available: https://doi.org/10.1038/s41598-021-94810-4 [22] D. B. Larremore, B. Wilder, E. Lester, S. Shehata, J. M. Burke, J. A. Hay, T. Milind, M. J. Mina, and R. Parker, “Test sensitivity is secondary to frequency and turnaround time for covid-19 surveillance,” medRxiv : the preprint server for health sciences, vol. 20, p. 20136309, Sep 2020. [Online]. Available: https://dx.doi.org/10.1101/2020.06.22.20136309 |
[23] CDC. (2020)
Research Use Only 2019-Novel Coronavirus (2019-nCoV) Real-time RT-PCR Primers
and Probes. Centers for Disease Control and Prevention. [Online]. Available: https://bit.ly/3EBvJyF [24] T. I. C. on
Non-Ionizing Radiation Protection, “Guidelines on limits of exposure to
ultraviolet radiation of wavelengths between 180 nm and 400 nm (incoherent
optical radiation),” Health Physics, vol. 87, no. 2, 2004. [Online].
Available: https: //doi.org/10.1097/00004032-200408000-00006 [25] M.
E. R. Darnell, K. Subbarao, S. M. Feinstone, and D. R. Taylor, “Inactivation of the
coronavirus that induces severe acute respiratory syndrome, SARS-CoV,” J Virol Methods, vol.
121, no. 1, pp. 85–91, Oct. 2004. [Online]. Available: https://doi.org/10.1016/j.jviromet.2004.06.006 [26] K. Bedell, A. H. Buchaklian, and S. Perlman, “Efficacy of an automated multiple emitter whole-room ultraviolet-c disinfection system against coronaviruses mhv and mers-cov,” Infection Control & Hospital Epidemiology, vol. 37, no. 5, pp. 598–599, 2016. [Online]. Available: https://doi.org/10.1017/ice.2015.348 [27] F. A. Alexandre, J. A. F. Faria,
and C. F. Cardoso, “Avaliação da eficiéncia da radiação
ultravioleta na esterilização de embalagens plásticas,” Ciéncia
e Agrotecnologia, vol. 32, no. 5, pp.
1524–1530, 2008. [Online].
Available: https://bit.ly/3mAbDie [28] J.
S. Smith, H. Hanseler, J. Welle,
R. Rattray, M. Campbell, T. Brotherton, T. Moudgil, T. F. Pack, K. Wegmann,
S. Jensen, J. Jin, C. B. Bifulco,
S. A. Prahl, B. A. Fox, and N. L. Stucky, “Effect of various decontamination procedures on
disposable n95 mask integrity and sars-cov-2 infectivity,” Journal of
clinical and translational science, vol. 5, no. 1, pp. e10–e10, Jun 2020.
[Online]. Available: https://dx.doi.org/10.1017/cts.2020.494 [29] M. Biasin, A. Bianco, G. Pareschi,
A. Cavalleri, C. Cavatorta,
C. Fenizia, P. Galli, L. Lessio,
M. Lualdi, E. Tombetti, A.
Ambrosi, E. M. A. Redaelli,
I. Saulle, D. Trabattoni,
A. Zanutta, and M. Clerici,
“Uv-c irradiation is highly effective in
inactivating sars-cov-2 replication,” Scientific Reports, vol. 11, no.
1, p. 6260, Mar 2021. [Online]. Available: https://doi.org/10.1038/s41598-021-85425-w [30] CDC. Implementing Filtering Facepiece Respirator (FFR) Reuse, Including Reuse after Decontamination, When There Are Known Shortages of N95 Respirators. Centers for Disease |
Control and
Prevention. [Online]. Available: https://bit.ly/3sD9ypQ [31] K. O’Hearn, S. Gertsman, M.
Sampson, R. Webster, A. Tsampalieros, R. Ng, J.
Gibson, A. T. Lobos, N. Acharya, A. Agarwal, S. Boggs, G. Chamberlain, E. Staykov, L. Sikora, and J. D.
McNally, “Decontaminating n95 and sn95 masks with ultraviolet germicidal
irradiation does not impair mask efficacy and safety,” vol. 106, no. 1, pp.
163–175. [Online]. Available: https://doi.org/10.1016/j.jhin.2020.07.014 [32] L.
Liao, W. Xiao, M. Zhao, X. Yu, H. Wang, Q. Wang, S. Chu, and Y. Cui, “Can n95
respirators be reused after disinfection? how many times?” ACS nano, vol. 14, no. 5, pp. 6348–6356, May 2020.
[Online]. Available: https://dx.doi.org/10.1021/acsnano.0c03597 [33] W. G. Lindsley, S. B. Martin, Jr, R. E. Thewlis, K. Sarkisian, J. O. Nwoko, K. R. Mead, and J. D. Noti, “Effects of ultraviolet germicidal irradiation (UVGI) on N95 respirator filtration performance and structural integrity,” J Occup Environ Hyg, vol. 12, no. 8, pp. 509–517, 2015. [Online]. Available: https://doi.org/10.1080/15459624.2015.1018518 [34] ABMT. 87 % das denúncias
de falta de EPIs são
sobre máscaras N95, diz AMB. Associação Brasileira
de Medicina do Trabalho.
[Online]. Available: https://bit.ly/3JlBFj5 [35] J.
Rose. (2020) Why can’t america make enough n95
masks? 6 months into pandemic, shortages persist. NPR. SPECIAL SERIES The
Coronavirus Crisis. [Online]. Available: https://n.pr/3mBjImT [36] K. Seresirikachorn, V. Phoophiboon,
T. Chobarporn, K. Tiankanon,
S. Aeumjaturapat, S. Chusakul,
and K. Snidvongs, “Decontamination and reuse of
surgical masks and N95 filtering facepiece
respirators during the COVID-19 pandemic: A systematic review,” Infect
Control Hosp Epidemiol,
vol. 42, no. 1, pp. 25–30, Jul. 2020. [Online]. Available: https://doi.org/10.1017/ice.2020.379 [37] E. C. Toomey, Y. Conway, C. Burton, S. Smith, M. Smalle, X.-H. S. Chan, A. Adisesh, S. Tanveer, L. Ross, I. Thomson, D. Devane, and T. Greenhalgh, “Extended use or reuse of single-use surgical masks and filtering face-piece respirators during the coronavirus disease 2019 (COVID-19) pandemic: A rapid systematic review,” Infect Control Hosp Epidemiol, vol. 42, no. 1, pp. 75–83, Oct. 2020. [Online]. Available: https://doi.org/10.1017/ice.2020.1243 |
[38] NEVMS, Descontaminação de máscaras de proteção respiratória modelo N95 (respirador particulado). Núcleo de Evidéncia de Mato Grosso do Sul, 2020. [Online]. Available: https://bit.ly/345EmW1 [39] B. K. Heimbuch, W. H. Wallace, K. Kinney, A. E. Lumley, C.-Y.
Wu, M.-H. Woo, and J. Wander, “A pandemic influenza preparedness study: use
of energetic methods to decontaminate filtering |
facepiece respirators contaminated with
H1N1 aerosols and droplets,” Am J Infect Control, vol. 39, no. 1, pp. e1–9,
Dec. 2010. [Online]. Available: https://doi.org/10.1016/j.ajic.2010.07.00 [40] S. Sangkham, “Face mask and medical waste disposal during the novel covid-19 pandemic in asia,” vol. 2, p. 100052. [Online]. Available: https://doi.org/10.1016/j.cscee.2020.100052
|